.TH comedi_calibrate 8 ""
|
||||
.SH NAME
|
||||
\fBcomedi_calibrate\fR - COMEDI calibration utility
|
||||
.SH SYNOPSIS
|
||||
\fBcomedi_calibrate\fR [-f /dev/comediN]
|
||||
.br
|
||||
.SH DESCRIPTION
|
||||
\fBcomedi_calibrate\fR is used to autocalibrate \fBcomedi\fR
|
||||
devices. Some \fBcomedi\fR devices have the ability to
|
||||
measure internal stable voltage references and control
|
||||
gains and offsets using programmable DACs. On these devices,
|
||||
\fBcomedi_calibrate\fR determines the proper values to
|
||||
feed into the programmable calibration
|
||||
DACs. These calibration settings are then
|
||||
saved to a file, which can be fed to comedi_apply_calibration()
|
||||
and related functions.
|
||||
|
||||
If comedi_calibrate finds an existing calibration file, it
|
||||
will not perform a new calibration by default. It will
|
||||
simply read its calibration information from the file.
|
||||
This behaviour
|
||||
may be overriden with the
|
||||
.B \-\-calibrate
|
||||
option, to force a new calibration to be performed.
|
||||
|
||||
After obtaining the calibration information for the board,
|
||||
comedi_calibrate will then apply a calibration setting
|
||||
appropriate for the specified subdevice, channel, and
|
||||
range (see the
|
||||
.B \-\-subdevice
|
||||
,
|
||||
.B \-\-channel
|
||||
, and
|
||||
.B \-\-range
|
||||
options below). See the comedilib documentation for
|
||||
comedi_apply_calibration() for more information.
|
||||
|
||||
\fBcomedi_calibrate\fR prints a lot of debugging information
|
||||
as it runs.
|
||||
|
||||
.SH LIMITATIONS
|
||||
|
||||
\fBcomedi_calibrate\fR only works on boards that it understands.
|
||||
For boards that are known to have autocalibration ability,
|
||||
but are not currently supported, \fBcomedi_calibrate\fR will
|
||||
output basic debugging information that is useful for
|
||||
the author to add support for your device. Please send this
|
||||
output to him.
|
||||
|
||||
.SH EXAMPLES
|
||||
|
||||
.TP
|
||||
.BI comedi_calibrate\ \-\-file\ /dev/comediN
|
||||
If no default calibration for \fI/dev/comediN\fR already exists,
|
||||
perform a calibration
|
||||
and save results to default location.
|
||||
|
||||
.TP
|
||||
.BI comedi_calibrate\ \-\-calibrate
|
||||
Perform a calibration, using existing calibration (if it exists) as
|
||||
starting point.
|
||||
|
||||
.TP
|
||||
.BI comedi_calibrate\ \-\-reset\ \-\-calibrate
|
||||
Perform a fresh calibration and save results to default location.
|
||||
|
||||
.TP
|
||||
.BI comedi_calibrate\ \-\-no\-calibrate\ \-\-subdevice\ subd\ \-\-range\ range\ \-\-save\-file\ cal_file
|
||||
Apply calibration for subdevice \fIsubd\fR, range \fIrange\fR, using
|
||||
the settings in the calibration file \fIcal_file\fR.
|
||||
|
||||
|
||||
.SH OPTIONS
|
||||
|
||||
\fBcomedi_calibrate\fR recognizes the following options:
|
||||
|
||||
.TP
|
||||
.BI \-a\ aref ,\ \-\-aref\ aref
|
||||
Along with \fB--channel\fR, \fB--range\fR
|
||||
and \fB--subdevice\fR, this specifies which calibration you want comedi_calibrate
|
||||
leave applied to the board. The default is 0.
|
||||
|
||||
.TP
|
||||
.B \-\-[no\-]calibrate
|
||||
Force calibration to be performed (or not).
|
||||
By default, comedi_calibrate will only perform a new calibration if
|
||||
no existing calibration file is found (see \fB--save-file\fR).
|
||||
|
||||
.TP
|
||||
.BI \-c\ channel, \-\-channel\ channel
|
||||
Along with \fB--subdevice\fR, \fB--range\fR
|
||||
and \fB--aref\fR, this specifies which calibration you want comedi_calibrate
|
||||
to leave applied to the board. The default is 0.
|
||||
|
||||
.TP
|
||||
.B \-\-[no\-]dump
|
||||
Perform (or not) dump of various information about
|
||||
state of calibration prior to performing new calibration.
|
||||
|
||||
.TP
|
||||
.BI \-f\ /dev/comediN, \-\-file\ /dev/comediN
|
||||
Perform calibration on device \fB/dev/comediN\fR.
|
||||
|
||||
.TP
|
||||
.B \-\-help
|
||||
Print help for options then exit.
|
||||
|
||||
.TP
|
||||
.B \-q, \-\-quiet
|
||||
Don't generate output to stdout.
|
||||
|
||||
.TP
|
||||
.BI \-r\ range ,\ \-\-range\ range
|
||||
Along with \fB--channel\fR, \fB--subdevice\fR
|
||||
and \fB--aref\fR, this specifies which calibration you want comedi_calibrate
|
||||
to leave applied to the board. The default is 0.
|
||||
|
||||
.TP
|
||||
.B \-\-[no\-]reset
|
||||
Reset (or don't reset) calibration subdevice to
|
||||
a reasonable starting point before
|
||||
performing calibration. The default is \fB--no-reset\fR. If
|
||||
\fB--no-reset\fR is used, the calibration routine will try to use
|
||||
an existing calibration (see \fB--save-file\fR) as a starting point.
|
||||
This is useful if
|
||||
a single calibration
|
||||
pass is not sufficient, and you wish to perform another calibration
|
||||
pass starting from the results of the previous pass.
|
||||
|
||||
.TP
|
||||
.B \-\-[no\-]results
|
||||
Perform (or not) dump of various information about
|
||||
state of calibration after performing new calibration.
|
||||
|
||||
.TP
|
||||
.BI \-S\ file ,\ \-\-save-file\ file
|
||||
Save calibration information to specified file,
|
||||
instead of default location. If the file already exists, comedi_calibrate
|
||||
will also use its contents as a starting point for the calibration pass (see
|
||||
\fB--no-reset\fR).
|
||||
|
||||
.TP
|
||||
.BI \-s\ subdevice ,\ \-\-subdevice\ subdevice
|
||||
Along with \fB--channel\fR, \fB--range\fR
|
||||
and \fB--aref\fR, this specifies which calibration you want comedi_calibrate
|
||||
to leave applied to the board. The default is 0.
|
||||
|
||||
.TP
|
||||
.B \-v, \-\-verbose
|
||||
Generate more verbose output to stdout, can be specified multiple times
|
||||
|
||||
|
||||
.SH SEE ALSO
|
||||
|
||||
\fBcomedi\fR(7)
|
||||
|
||||
.SH VERSION
|
||||
|
||||
0.7.x
|
||||
|
||||
.SH AUTHOR
|
||||
|
||||
David Schleef, <ds@schleef.org>
|
||||
|
||||