I always seem to forget how to convert the calibration
dump information into code for doing a calibration, so
I'm writing this mostly for myself.

Boards may have one of 4 calibrations statuses, depending
on how well the calibration code is trusted.  These are:
STATUS_NONE, the default for no information; STATUS_SOME,
meaning that a dump has been converted to initial code,
but not tested; STATUS_DONE means that the output of a
STATUS_SOME dump has been checked, and is correct;
STATUS_GUESS is a marker that code has been converted
from a previous version of the code, but not checked.

The NI E series boards have several internal voltages that
can be measured, and also several calibration DACs that function
similar to adjustable resistors on old data acquisition boards.
The information we need is which DACs affect which measurable
voltages; then we can write calibration code that adjusts those
DACs until the voltages are within spec.

Usually, there are DACs (or multiple DACs) that are added to
an analog input signal: 1) before the variable gain amplifier
("pre-gain"), 2) after the variable gain amplifier ("post-gain"),
3) between the board's stable voltage reference and the
reference input to the ADC ("gain offset"), and 4) before a
unipolar-to-bipolar adjuster ("unipolar offset"), or other
equivalent circuit.

In addition there are DACs that adjust the output voltages and/or
reference voltage inputs to a D/A converter.  These are pretty
intuitive once analog input is understood, and is dependent on
correct analog input calibration.

The measurable quantities are 0 volts and an internal voltage
reference near 5 volts, and can be measured at any gain.  The
interesting combinations are:

  ai, bipolar zero offset, low gain
  ai, bipolar zero offset, high gain
  ai, bipolar voltage reference, low gain
  ai, unipolar zero offset, low gain

The unipolar zero offset may not be available on some boards.

In a STATUS_NONE dump, for each measurable quantity and each
calibration DAC, the DAC is varied throughout its entire range
and the quantity measured.  The data is linearly fit, and if
the slope is statistically non-zero, a line is printed:

  caldac[0] gain=1.26(11)e-7 V/bit S_min=235.659 dof=254

The information given is caldac index, slope (gain) and slope
error (in parenthesis, modifying the last two digits of the
slope), and two statistical parameters S_min and degrees
of freedom.  S_min and dof will be roughly similar for a
good fit.  If S_min is more than a factor of 4 greater than
dof, this is probably not a good fit.  Typically this means
that the DAC doesn't affect the measureable strictly linearly,
or there is systematic noise.  The latter seems to common in
E series boards, so I'm not too worried about the following
dump where there are S_min/dof ratios above 4.

Here's an example dump, generated by a STATUS_NONE dump for
a pci-mio-16xe-10, with the analog output section removed:

  Warning: device not fully calibrated due to insufficient information
  Please send this output to <ds@schleef.org>
  Id: comedi_calibrate.c,v 1.21 2001/10/10 22:07:53 ds Exp
  Driver name: ni_pcimio
  Device name: pci-mio-16xe-10
  Comedi version: 0.7.61
  ai, bipolar zero offset, low gain
  offset -6.795(14)e-3, target 0
  caldac[0] gain=1.26(11)e-7 V/bit S_min=235.659 dof=254
  caldac[2] gain=3.96840(14)e-4 V/bit S_min=1390.18 dof=254
  caldac[3] gain=4.348(11)e-6 V/bit S_min=258.75 dof=254
  caldac[8] gain=5.4659(69)e-7 V/bit S_min=386.361 dof=254
  ai, bipolar zero offset, high gain
  offset -2.4224(55)e-4, target 0
  caldac[0] gain=3.61(45)e-9 V/bit S_min=247.26 dof=254
  caldac[2] gain=3.96644(48)e-6 V/bit S_min=351.927 dof=254
  caldac[3] gain=4.063(46)e-8 V/bit S_min=272.024 dof=254
  caldac[8] gain=5.46305(30)e-7 V/bit S_min=314.035 dof=254
  ai, bipolar voltage reference, low gain
  offset 4.992959(13), target 5
  caldac[0] gain=-4.4928(11)e-5 V/bit S_min=1111.4 dof=254
  caldac[1] gain=-2.792(11)e-6 V/bit S_min=248.971 dof=254
  caldac[2] gain=3.96488(14)e-4 V/bit S_min=1059.18 dof=254
  caldac[3] gain=4.318(11)e-6 V/bit S_min=437.441 dof=254
  caldac[8] gain=5.4810(70)e-7 V/bit S_min=404.213 dof=254
  ai, unipolar zero offset, low gain
  offset nan, target 0
  caldac[2] gain=3.96773(39)e-4 V/bit S_min=158.236 dof=107

[The explanation gets a little fuzzy here]

The resulting function for calibration will look something like:

  void cal_ni_pci_mio_16xe_10(void)
  {
	postgain_cal(ni_zero_offset_low, ni_zero_offset_high, XXX);
	cal1(ni_zero_offset_high, XXX);
	cal1(ni_reference_low, XXX);
	cal1(ni_unip_offset_low, XXX);
  }

You get to fill in the XXX's.  The post-gain calibration DAC will
be the one for which the ratio of caldac slopes for the low and
high gain measurables is similar to the ratio of input ranges for
low and high gain.  This ratio is typically 100 or 200, and really
should be printed by the program.  Thus, for this dump, we choose
caldac[2], since the ratio is very nearly 100.  We don't choose
caldac[0] or caldac[3], because the gains are smaller, and the
ratio isn't exactly 100 or 200.

Next is the pre-gain calibration.  Adding a voltage before the
amplifier will affect every input range selection equally, so the
pre-gain cadac slope will be nearly equal for both bipolar zero
offset at low and high gain.  In this example, it would be caldac[8].

Next is the voltage reference calibration.  The caldac controlling
the voltage reference adjustment is proportional to the offset,
so the correct caldac will typically be the one that has a large
slope for the bipolar voltage reference measurement, but a small
slope (by a factor of 2e4, here) for the zero offset measurements.
It could be any of caldac[0], caldac[1], or caldac[3], or possibly
all of them.  We'll choose the caldac with the largest slope for
rough calibration, then use the one with the smallest slope for
fine calibration, namely caldac[0] and caldac[1].

This is one way that STATUS_SOME is useful, because you can calibrate
the zero offset, then get a much better idea which other channels
are likely to be for the voltage reference.

In this example, there doesn't appear to be a caldac that affects
unipolar zero offset, so it will not be used in the final function:

  void cal_ni_pci_mio_16xe_10(void)
  {
	postgain_cal(ni_zero_offset_low, ni_zero_offset_high, 2);
	cal1(ni_zero_offset_high, 8);
	cal1(ni_reference_low, 0);
	cal1(ni_reference_low, 1);
  }